logo

szczegółowe informacje o produktach

Created with Pixso. Do domu Created with Pixso. produkty Created with Pixso.
Stacjonarne urządzenia badawcze liczników
Created with Pixso.

Flexible High-Precision Modular Design Energy Meter Testing Equipment for Accurate Power Measurement

Flexible High-Precision Modular Design Energy Meter Testing Equipment for Accurate Power Measurement

Nazwa marki: GENY
Numer modelu: YC-1891D
MOQ: 1 komputery
Cena £: Zbywalny
Warunki płatności: L/C, T/T, D/P.
Zdolność do zaopatrzenia: 100/rok
Szczegółowe informacje
Miejsce pochodzenia:
CHINY
Orzecznictwo:
ISO, CE
Wyjście napięcia testowego (neutralne fazowe):
24 V - 300 V.
Moc bieżącego wyjścia:
1500VA
Zakres częstotliwości:
45 Hz 65 Hz
Moc wyjścia napięcia:
2000va
Szczegóły pakowania:
Sprawa sklejka
Możliwość Supply:
100/rok
Podkreślić:

Flexible Energy Meter Testing Equipment

,

High-Precision Meter Test Equipment

,

Modular Design Stationary Meter Test Equipment

Opis produktu
Flexible Energy Meter Testing Equipment
GENY's Meter Test Equipment utilizes advanced electronic measurement technology and a modular design. By integrating differential reference standards with high-precision power amplifiers, our systems deliver accurate and reliable performance.
Ideal Applications
  • Electric energy meter manufacturers
  • Electric power utilities
  • Metrology laboratories
  • Measurement and quality inspection departments
  • Industrial factories and enterprises
Our solutions are designed to meet the demanding requirements of modern power measurement and testing environments.
Technical Parameters
Power Supply
Voltage 3×230/398V±10%
Frequency 50/60Hz±2Hz
Power capacity > 6kVA
Ambient temperature -10°C -- +40°C
Relative humidity 35% -- 85%
Power Source
Voltage
Test voltage output (Phase-Neutral) 24V - 300V
Power of the voltage output 1000VA
Resolution Better than 0.01% of full scale value of range
Setting accuracy Better than 0.05% at the final range value
Stability Better than 0.005%/h (integration time 150 s)
Load regulation from 0-max. load < 0.01%
Distortion factor < 0.3% for linear resistance load
Harmonics 2nd - 21st free programmable
Current
Test current output 1mA - 120A
Power of the current output 1500VA
Resolution Better than 0.01% of full scale value of range
Setting accuracy Better than 0.05% at the final range value
Stability Better than 0.005%/h (integration time 150 s)
Load regulation from 0-max. load < 0.01%
Distortion factor < 0.3% for linear resistance load
Harmonics 2nd - 21st free programmable
Phase angle
Range 0 -- 360°
Resolution 0.01°
Setting accuracy 0.1°
Frequency
Range 45Hz - 65Hz
Resolution 0.01Hz
MSVT
Number of secondary 50
Voltage range 300V
Ratio 1:1
Primary to secondary ratio error 0.25%
Secondary to secondary ratio error 0.02%
Secondary to secondary Phase error 0.02°
Frequency range 45Hz - 65Hz
Output power (per winding) 15VA
Flexible High-Precision Modular Design Energy Meter Testing Equipment for Accurate Power Measurement 0 Flexible High-Precision Modular Design Energy Meter Testing Equipment for Accurate Power Measurement 1 Flexible High-Precision Modular Design Energy Meter Testing Equipment for Accurate Power Measurement 2